Oman Ministry of Health Circular No. 160/2025
Regarding Recall of Continuous Glucose Monitoring System from Dexcom
To: All Medical Device Establishment
Subject: Regarding Recall of Continuous Glucose Monitoring System from Dexcom
After Compliments,
Recall of Continuous Glucose Monitoring System from Dexcom
Device Name: Continuous Glucose Monitoring System from Dexcom
Manufacturer: Dexcom Inc
Local agent: Scientific Pharmacy
Affected devices:
STK-GT-001 lot 19492808
STK-D7-014 lot 20280910
STK-D7-014 lot 20280914
Reason:
A manufacturing defect has been identified in the foam used in the production of some devices, which may result in the absence of audible alerts in the affected devices mentioned above.
Instructions:
Immediately stop using the affected devices.
Contact the authorized distributor to replace the affected devices.
Conduct regular testing of the speaker to ensure the audio alerts are functioning properly and there is no defect.